VELP UDK Distillation Units

VELP Scientifica UDK Distillation Units are the ideal solution for performing analyses concerning different applications such as determining ammoniacal nitrogen, protein nitrogen, (Kjeldahl or direct alkaline distillation), nitric nitrogen (after reduction), phenols, volatile acids, cyanides, alcohol content, Sulphur dioxide, TVBN, ad Devarda nitrogen determination.

These units support the most advanced technology, consisting in a unique patented steam generator and an outstanding efficient patented titanium condenser that are wisely combined with a technopolymer splash head. Designed with a strong and chemical-resistant structure made of technopolymer, UDK Series has been designed to last in time and to perform reliable analysis for many years. UDK series also incorporates TEMS™ technology for major savings in time, energy, money, and space pursuing VELP’s contribution to environmental protection.

Choose from UDK 129 or 139 Series:

The UDK 129 Series runs automatically, after setting sodium hydroxide addition and distillation time using the LCD display to get reliable and accurate results. The high-precision pumps ensure constant accurate dosing of reagents and the cooling water is automatically stopped during pauses, thus cutting down on its consumption. The UDK 129 incorporates the same high-level technology as the top of the range, with the VELP patented steam generator that offers high performance, safety (no pressure inside) and is maintenance-free. Another unique VELP component is the titanium condenser offering reduced water consumption, ensuring that distillate temperature always remains below the threshold value. The unit works with a technopolymer splash head that ensures durability to protect your investment and requires no maintenance. The technopolymer housing ensures high resistance to chemicals and long life. The UDK 129 had numerous safety features to provide maximum protection for the user. Continuous monitoring indicates incorrect tube and handle positioning; the cooling water flow detector provides a high lever of safety. With a novel design, a level is used to displace the tube support enabling sample tubes to be inserted without any effort and clamped in place securely. Technologically advanced, the UDK 129 includes many features that ensure efficient and reliable distillation, far beyond expectations of an ordinary entry level unit.

The UDK 139 Series runs automatically, after setting distillation time, water, and sodium hydroxide addition and steam generation output level between 10 and 100% using the innovative 3.5” color touchscreen. The high-precision pumps ensure constant accurate dosing of reagents. Accessing the 10 customizable methods available in 6 different languages is simple and intuitive. The UDK 139 incorporates a considerably high level of technology, with the VELP patented steam generator that offers high performance, safety (no pressure inside) and is maintenance-free. Another unique VELP component is the titanium condenser offering reduced water consumption, ensuring that distillate temperature always remains below the threshold value.

Key Features

  • Safety level for users including a safety lever which avoids contact with soiled surfaces
  • Protective doors with sensor shield test tube and prevents spills
  • Service door + automatic electrical shutdown for extraordinary maintenance
  • Cooling water flow-rate detector activates low flow-rate warning signal
  • Test tube sensor that ensures the presences of the test tube
  • Drip tray to collect any drops

Specifications for VELP UDK Distillation Units
  UDK 129 UDK 139
Structure: Corrosion-resistant technopolymer  
Display: LCD 3.5" color touch screen
Time related to start for
Devarda alloy analysis:
0 - 99 min
Protocol library: 1 customizable method 10 customizable methods
Reproducibility (RSD): = 1 %  
Recovery: = 99.5 % at nitrogen levels between 1 and 200 mg N
Detection Limit: = 0.1 mg N
Power: 2100 W
Weight: 55 lb 57.3 lb
WxHxD: 15.2" x 30.7" x 16.4"
PERFORMANCE
Distillation time: 5 minutes to collect 100 ml of distillate 4 minutes to collect 100 ml of distillate
Sodium hydroxide addition: automatic
Sodium hydroxide volume: 0 - 100 ml 0 - 150 ml
Tap water consumption: from 0,5 l/min at 15°C - from 1 l/min at 30°C 0 - 200 ml
Distillation time: settable
Distillation residues removal: automatic
Steam flow regulation: from 10 to 100 %
Tap water consumption: from 0,5 l/min at 15 °C - from 1 l/min at 30 °C
Connections: 2 x USB for printer, mouse
Filter
Item Model No. Power Unit of Measure Price Purchase
VLP794001083
F30200120
UDK 129 230V/50-60 Hz Each
$14,054.95
Contact customer service for ETA In Stock
VLP794001085
F30200130
UDK 139 230V/50-60 Hz Each
$11,162.95
Contact customer service for ETA In Stock

With Selected Items: