Glove Boxes

Hypoxia Chamber Glove Box

The hypoxia chamber glove box is ideal for tissue culture work, including tumor cell and stem cell research. It features a color operated “touch screen” for automatically controlling oxygen levels, temperature levels, and carbon dioxide levels.
Basic features include:
Hypoxia Chamber Glove Box
The hypoxia chamber glove box is ideal for tissue culture work, including tumor cell and stem cell research. It features a color operated “touch screen” for automatically controlling oxygen levels, temperature levels, and carbon dioxide levels.
Basic features include:
- 8” color panel touch screen
- O2 sensor and CO2 sensor
- Platinum RTD with sealing gland
- N2, CO2, and O2 gas hook-up
- Ethernet port connection for data logging
- Thermoelectric Assembly (Peltier device)
- Password protected, configurable PID control
- One touch, on screen data logging
- Air circulation fan
- Configurable for up to six individually controlled sample chambers
Product Specifications:
| Oxygen (O2) Range | 0-25% |
| Oxygen sensor accuracy | ±0.25% |
| Carbon Dioxide (CO2) range | ± 0-20% C |
| Carbon Dioxide sensor accuracy | ± 0.2% |
| Temperature range | Ambient to 40° C |
| Temperature Accuracy | ± 0.3 °C @ 0 °C (class “B” accuracy) |
| Parameter control: | ± 1% or degree of set point |
| Outside Dimensions | 55” W x 35” D x 38” H |
| Inside Dimensions | 41” W x 28” D x 26” H |
| Power Requirements | 115 V 6.5 Amps/60 Hz, 230V 5 Amps/50 Hz |